2.Edit Exposure Mode Settings

2.4.2 Setting low-density sections

1
Enable dynamic range adjustment (BRIGHT REGION)
Each time [BRIGHT REGION] is touched, the setting switches between the highlight
(white outline characters against a colored background) and normal display. This is
enabled when in the highlighted state.
Screen Appearance when Dynamic Range Adjustment (BRIGHT REGION) is Enabled
2
Set the minimum density
Touch the
or
20.
The value on the left is the setting saved previously, and the value on the right is the
modified value.
This value determines the minimum density assigned to the subject range. Larger values
result in higher minimum densities, and smaller values result in lower minimum densities.
NOTE:
If the EFFECT parameter is set to a value that is too high, artifacts can occur
in the image, or the effective data of the exposure image may be damaged.
Check the exposure image, and if the image is proccessed excessively, either
set to a smaller minimum density or disable the setting.
button, and set the EFFECT parameter within the range between 1 to