Baumer 11228779 Betriebshandbuch - Seite 14
Blättern Sie online oder laden Sie pdf Betriebshandbuch für Zubehör Baumer 11228779 herunter. Baumer 11228779 29 Seiten. Off-highway ground & crop radar sensor
Manual R600V.DAE0 (EN)
Strong re-
flections
Fig 7
Shifting reflections in a stationary situation
2.2.3
On Vehicle Calibration
In a stationary situation (vehicle is not moving) while measuring on structured surfaces (such
as soil) the output distances may shift over time, due to slight changes in position due to
vibration and varying radar wave interference. For on-machine calibration purposes a flat sur-
face of adequate size (e.g. larger than 1200mm by 1200mm) with good radar reflection, and
with its geometric centre lined up with the sensor centre axis must be used to avoid this effect
and ensure a stable calibration procedure.
NOTE
For distance calibration on a vehicle a defined target must be used.
Page 14 of 29
Directionality
(opening angle)
Distance averaging
(over opening angle)
R600V.DAE0_manual_1.5 (EN).docm
Last Update: 07 Jan 2021