Apogee SQ-100X Series Benutzerhandbuch - Seite 6

Blättern Sie online oder laden Sie pdf Benutzerhandbuch für Zubehör Apogee SQ-100X Series herunter. Apogee SQ-100X Series 19 Seiten. Quantum sensor

SPECIFICATIONS

Power Supply
Sensitivity
Calibration Factor
(Reciprocal of Sensitivity)
Calibration Uncertainty
Calibrated Output Range
Measurement
Repeatability
Long-term Drift
(Non-stability)
Non-linearity
Response Time
Field of View
Spectral Range
Directional (Cosine)
Response
Temperature Response
Operating Environment
Dimensions
Mass
Cable
Calibration Traceability
Apogee SQ series quantum sensors are calibrated through side-by-side comparison to the mean of transfer
standard quantum sensors under a reference lamp. The reference quantum sensors are recalibrated with a 200 W
quartz halogen lamp traceable to the National Institute of Standards and Technology (NIST).
± 5 % (see Calibration Traceability below)
Less than 1 % (up to 4000 µmol m
370 to 650 nm (wavelengths where response is greater than 50 % of maximum; see Spectral
± 5 % at 75° zenith angle (see Cosine Response below)
10 to 60 C; 0 to 100 % relative humidity; can be submerged in water up to depths of 30 m
5 m of two conductor, shielded, twisted-pair wire; TPR jacket (high water resistance, high UV
stability, flexibility in cold conditions); pigtail lead wires; stainless steel (316), M8 connector
SQ-100X-SS
Self-powered
-2
-1
0.1 mV per µmol m
s
-2
-1
10 µmol m
s
per mV
0 to 400 mV
Less than 0.5 %
Less than 2 % per year
Less than 1 ms
180°
Response below)
- 0.04 % per C
24 mm diameter, 33 mm height
90 g (with 5m of lead wire)
located 25 cm from sensor head
-2
-1
s
)