Setting the weight unit
Select the weight unit (kg/oz/lb) by pressing the Units key. The corresponding LCD status indicator
(16, 17 or 18) for the current weight unit lights up.
Determining the weight
Press the ON key for at least 3 seconds to turn the scales on. An automatic function test is carried out
after being switched on. This is completed by displaying zero. The scales are now ready to operate. If
the display shows zero, the scales can be loaded. The weight can be read directly after a period of
standstill.
Resetting the scales
Wait until the standstill status indicator
(14) is on. By pressing the ZERO key (8), measuring inac-
curacies are compensated for after the scales are switched on and reset before the actual weighing
process. The status indicator
(6) is on when the scales are reset.
Tare functions
Taring the scales:
Use the TARE key (9) for example if the patient is being weighed with other items (e.g. clothing, shoes
or other items). To do this, weigh these items first. After the weight is displayed, press the TARE key
(9) to reset the display again. The ADE-MAK changes to net mode (NET status indicator (5) is on).
The weight of the previously weighed and saved items is deducted later when weighing the patient
and only the patient weight is displayed.
Deleting saved tare values:
Remove the weight from the scales and wait until the standstill status indicator
(14) lights up.
Press the TARE key (9). The ADE-MAK changes to gross mode (NET status indicator (5) goes out).
Enter the manual tare value:
Enter the appropriate tare value using the key pad. If no value is entered "ER TOUT" will appear on
the display. The ADE-MAK changes automatically to weight mode. If the tare value has been entered
correctly, the value will be saved in the unit. Press the ENTER key (or TARE). The ADE-MAK is now
ready for weighing and changes to net mode [NET status indicator (5) is on].
The manual tare value must be >0, the maximum value is limited by the weighing range.
PRESET TARA adjusted:
Press and hold the PRESET TARE key (10) [approx. 3 sec.], PRO PT X appears on the display. Then
you can select the memory location (e.g. 2). Now enter the TARE VALUE and confirm with ENTER.
PRESET TARA
Accessing the preset
value:
Press the PRESET TARE key (10). "Pt x" appears on the display. Enter a number from 0-9 (memory
location for the tare value). If no number is entered within five seconds, the unit will change back to
weight mode.
After entering the number, the ADE-MAK accesses the preset tare value, shows it for five seconds on
the display and then changes to net mode. The accessed tare value is accepted at the same time.
PRESET TARA
Changing the preset
value:
Press the PRESET TARE key (10) for two seconds to change a preset tare value. PRO PT X will ap-
pear on the display.
Select a number from 0-9 (memory location for the tare value).
The ADE-MAK accesses the preset tare value at this memory location and shows it on the display.
Enter the new value and press the ENTER key. The unit saves the new value at the selected memory
location.