A-M Systems 4000 Gebrauchsanweisung
Box ID
This 8 position slide switch assigns the unit identifier to each
Model 4000 in your system. If you have only a single Model 4000,
this value should be set to "1". If you have more then one Model
4000, then each unit should have a unique number.
The BOX ID:will determine the channel assignments recorded by the Model 4000 as seen by the
Datawave Sciworks System.
.
Control for Additional Units
These DB25 connectors enable multiple units to be controlled by a single
Datawave Sciworks System. These connectors are equipotent, i.e., either can
serve as an input/output to any other Model 4000 unit.
TBSI
M-SERIES
HEADSTAGE
CHANNELS
BOX ID
1
1-30
2
31-60
3
61-90
4
91-120
5
121-150
6
151-180
7
181-210
8
NA
Figure 6.
TBSI
T-SERIES
HEADSTAGE
CHANNELS
1-32
33-64
65-96
97-128
129-160
161-192
193-224
125-256
Box ID
Box ID
Figure 7.
Control for
Additional Units
Control for
Additional Units
8