Aqua Lab PRE Schnellstart-Handbuch - Seite 5

Blättern Sie online oder laden Sie pdf Schnellstart-Handbuch für Messgeräte Aqua Lab PRE herunter. Aqua Lab PRE 17 Seiten. Water activity meter

Aqua Lab PRE Schnellstart-Handbuch

Predictive Reading

A predictive reading shows on the screen after about
40 seconds. Reading is complete when your instrument
'beeps' and the LED blinks. Instrument then displays the
sample's final water activity reading and temperature.

Final Primary Standard Reading

Take two readings. The water activity readings should be
within +/- 0.01 water activity of the given value for the
verification standard. If you are using the 0.760 standard
as your primary standard, it should measure between 0.75
and 0.77*. If your primary standard measures in spec,
move on and determine your second verification standard.
If not, clean the instrument and retry the standard. Dispose
of sample cup.
*Accuracy of the AquaLab Pre is ±0.01 water activity.
Note: If you are using an instrument with a volatiles block because of volatiles
in your products, your instrument is in spec if it measures between 0.745 and
0.775. The volatiles sensor for the Pre measures to ±0.015 water activity.
4