3.3.
ORP (R65) 4-20 mA Probes
1. ORP 4-20 mA probes have four wires; black, red, green and white. The red wire is to be connected
to the +24 VDC terminal and the back wire to the 24 VDC common terminal via the load in the loop.
2. For an instrument with a range of 0 to 1000 mV the green and white wires are to be shorted.
3. For an instrument with a range of –500 to 500 mV the green and white wires are to be isolated from
each other.
3.4.
Submersion Mounting Series 8 Differential Probes
1. Refer to Figure 5.
2. A submersion mounting kit, STC60L is available from Water Analytics which includes 4 ft. of 1"
pipe, 1-1/2" x 1" reducer, a strain relief fitting and wire mounting bracket. Proceed as follows, either
with the kit or with your own hardware.
3. Apply a thread sealant to the thread on the cable end of the probe and screw a 1-1/2" x 1" NPT
reducer onto the probe. Route the sensor cable through an appropriate length of 1" pipe and using
thread sealant, screw the pipe into the reducer. The cable end of the probe should not be exposed
to the process. A cable strain relief fitting should be used on the upper end of the pipe. In the kits a
wire bracket is provided to aid in supporting the assembly.
NOTE: An optional protective shroud, Part No. PROTECTOR-3 should be used on the electrode
end of the probe to protect the electrodes from accidental contact with the tank bottom, sides or
objects in the process.
3.5.
Submersion Mounting Series 6 Differential Probes
1. Refer to Figure 4.
2. A submersion mounting kit, STC60-6 is available from Water Analytics which includes 4 ft. of 1"
pipe, 1-1/2" x 1" reducer and a strain relief fitting. Proceed as follows, either with the kit or with your
own hardware:
3. Install the optional protective shroud, Part No. PROTECTOR-6 on the probe by threading the probe
cable through it. The shroud will contact the shoulder on the probe.
4. Install the compression fitting components on the probe in the order shown in the drawing below so
that the pipe thread is towards the cable end of the probe. If you are concerned that the shroud
may get pushed up and expose the electrodes you can lock it down by the positioning of the fittings.
5. Snug up the nut of the compression fitting to locate it in the desired position. Hand tighten as much
as possible, then turn 1/2 turn with a wrench.
6. Apply a thread sealant to the pipe thread portion of the compression fitting and screw a 1-1/4" x 1"
NPT reducer to it.
7. Route the sensor cable through an appropriate length of 1" pipe and using thread sealant, screw the
pipe into the reducer on the probe. The cable end of the probe should not be exposed to the
process.
3.6.
Flow-through tee mounting Series 8 Differential Probes
1. Refer to Figure 5.
2. Apply pipe sealant to the electrode end of the probe and screw it into the AquaMetrix union tee w/
adaptor (Part No. AM-MH538N9A) or any standard 1-1/2" NPT tee.