3A HEALTH CARE MINIASPEED BATTERY PRO Gebrauchsanweisung - Seite 7
Blättern Sie online oder laden Sie pdf Gebrauchsanweisung für Medizinische Ausrüstung 3A HEALTH CARE MINIASPEED BATTERY PRO herunter. 3A HEALTH CARE MINIASPEED BATTERY PRO 10 Seiten. Professional aspirator
Auch für 3A HEALTH CARE MINIASPEED BATTERY PRO: Gebrauchsanweisung (10 seiten), Gebrauchsanweisung (10 seiten)
3.2 Secretion collection vessel with single use bag (8.):
The aspirator can be used with the 1000 cc re-usable transparent secretion collection vessel (2.) and with the single use bag (8.)
supplied. In this case the antibacterial filter is integrated in the single-use bag, therefore the antibacterial filter (3.) and the blue lid with
the valve (2a.) should not be used. The filter embedded in the bag, also prevents the reflux of the liquids sucked towards the aspirator
when it is full, or when it is inadvertently turned over. In this case to restore the device to normal operation, the single use
bag shall be replaced. For the cleaning and disinfecting operations of the tubes (4. and 5.) and vessel (2.), sterilise the single
parts in autoclave at a maximum temperature of 121°C, or by boiling them for 10 minutes. The bag is single use and it MUST be
replaced after each use. The bag must be completely inserted in the vessel in order to prevent any vacuum losses.
N.B.: maximum disposable bag usage vacuum: -0.75 bar (75 kPa).
3.2.1 Connection: connect one end of the short sterilisable silicon tube (4.) to the yellow rubber holder (VACUUM) of the bag (8.)
and the other end to the "INLET" input (1e.) of the aspirator. Connect one end of the long sterilisable silicon tube (5.) to the red rubber
holder (PATIENT) of the bag (8.) and connect the sterile single-use manual regulator (6.) and the single-use sterile cannula (7.) to the
other end.
N.B.: only use the single use bags supplied by 3A - Code 3AC461.
N.B.: If using chemical disinfectants, carefully follow the manufacturer's instructions.
• The cannula and the manual regulator are sterile, disposable products and must be replaced after every application.
• The disposable antibacterial filter must be replaced after every use.
• Never wash the device under running water or by immersion. Clean the outside of the device using only a cloth dampened with a non-
abrasive detergent.
TECHNICAL SPECIFICATIONS OF THE ACCESSORIES:
• Disposable antibacterial filter - code 3A1385
• 1000 ml collection vessel in polycarbonate, complete with cap - code 3AC286
• Silicone tube Ø 6 x 12 mm L 240 mm - code 3A476
• Silicone tube Ø 6 x 12 mm L 1300 mm - code 3A561
• Disposable sterile cannula CH14 - code 3A4167
• Disposable manual regulator - code 3A560
• 1 litre disposable bag - code 3AC461
PROBLEMS
Excessive noise.
The unit switches on but does not aspirate
The vacuum rate cannot be regulated
When the appliance is switched on, the
protection fuse always trips
The vacuum gauge does not work
Note: if you experience faults or malfunctioning problems different to those listed above, always and exclusively
contact authorised assistance centres.
CLEANING AND DISINFECTION PROCEDURES
PROBLEMS, CAUSES and SOLUTIONS
POSSIBLE CAUSES
Damaged pump or blockages in the internal
aspiration circuit
- Damaged pump
- Vacuum regulator fully open.
Connection tubes disconnected and/or badly
connected, broken connection tubes. Contai-
ner not in a vertical position, full, or defective
overflow valve. Possible blockage of
the hydraulic circuit inside the unit.
Damage to the internal hydraulic system or
blockage of the connection tubes to the aspi-
ration unit.
Pump probably damaged or in shortcircuit.
Liquid penetrating the pneumatic circuit.
Send to the assistance circuit
- Send to the assistance circuit.
- Check the position of the vacuum regula-
tor. Check the connections and the integri-
ty of the tubes. Position the container in a
vertical position, check the overflow valve
(blocked) and/or replace the silicon tubes.
Send to the assistance circuit
Send to the assistance circuit
Send to the assistance circuit
11
SOLUTIONS