Olympus GX41 Anleitung Handbuch - Seite 23
Blättern Sie online oder laden Sie pdf Anleitung Handbuch für Mikroskop Olympus GX41 herunter. Olympus GX41 32 Seiten. Compact inverted metallurgical microscope
SPECIFICATIONS
Item
1. Optical system
2. Illumination system
3. Focusing mechanism
4. Revolving nosepiece
5. Observation
Binocular tubes
tubes
Trinocular tubes
Tilting tubes
6. Eyepieces
7. Objectives
8. Stage
9. Mechanical stage
10. Dimensions and weight
11. Operating Environment
UIS2 (UIS) (Universal Infinity System) optics
6V30WHAL-L long-life halogen bulb.
Hosobuchi G4 20H CF-6.
(Average bulb life: Approx. 2000 hrs.
under rated use)
Output rating: 6 V DC, 30 VA
Input rating: 100-120/220-240 V
Vertical movement of revolving nosepiece (Stage is fixed.)
Coarse/fine adjustment knobs (provided with tension adjustment mechanism).
Stroke (from focal point on stage surface): 7 mm upward and 2 mm downward.
Quadruple-position revolving nosepiece (Fixed)
U-CBI30-2: Field number 20, tube
inclination 30°, interpupillary distance
adjustment range 48-75 mm.
U-CTR30-2: Field number 20, tube
inclination 30°, interpupillary distance
adjustment range 48-75 mm, light
intensity ratio between binocular and
straight tubes fixed (at 50:50).
U-CTBI: Field number 18, tube inclination
30-60°, interpupillary distance adjustment
range 48-75 mm. 10X eyepieces are fixed
and not exchangeable.
CKX-TBI: Field number 20, tube
inclination 30-60°, interpupillary distance
adjustment range 50-76 mm. WHB10X
eyepieces are provided and not
exchangeable.
WHB10X, WHB10X-H:
Field number 20. Eyepiece micrometer
disk acceptable.
MPlan5X/10X/20X/40X/50X/100X.
Stage dimensions 160(W) x 250(D) mm
Transversing area: 120(X) x 76(Y) mm.
Coaxial low drive control knobs attachable on right or left side of plain stage.
Provided with stage plate holder, slide glass holder and stage plate.
236(W) x 624(D) x 407(H) mm (when a binocular tube is mounted). Approx. 10 kg
(22 lbs).
· Indoor use.
· Altitude: Max. 2000 meters.
· Ambient temperature:5° to 40°C (41° to 104° F).
· Relative humidity:85% for temperatures up to 31°C (88°F), decreasing linearly through
70% at 34°C (93°F), 60% at 37°C (99°F) to 50% relative humidity at 40°C (104°F).
· Supply voltage fluctuations: ±10%.
· Pollution degree: 2 (in accordance with IEC60664).
· Installation (overvoltage) category: II (in accordance with IEC60664)
Specifications
6V30WHAL high-intensity halogen bulb.
Philips Type 5761.
(Average bulb life: Approx. 100 hrs.
under rated use)
AC, 0.85/0.45 A, 50/60 Hz.
U-BI30-2: Field number 22, tube
inclination 30°, interpupillary distance
adjustment range 50-76 mm.
U-TR30-2: Field number 22, tube
inclination 30°, interpupillary distance
adjustment range 50-76 mm, light
intensity ratio between binocular and
straight tubes variable in 3 steps (100:0,
20:80, 0:100).
U-TBI-3: Field number 22, tube inclination
5-35°, interpupillary distance adjustment
range 50-76 mm.
Only WH10X eyepieces can be used.
WHN10X, WHN10X-H:
Field number 22. Eyepiece micrometer
disk acceptable.
GX41