VIRALERT
3

3 SYSTEM SETUP

3.1 Measurement Location

1)
Choose a location near to the entrance of the area you want to protect.
2)
There must be space to install the Camera and Blackbody Reference Source
so that the Measurement Subject can face the Camera from a distance of
1 m / 3 ft.
3)
Avoid locations where there are possible heat sources e.g. electrical
equipment, radiators etc. in the field of view of the camera.
4)
Also avoid locations where the camera may pick up reflections from windows,
mirrors etc.
5)
During the temperature measurement process, the Subject will need to stand
still in front of the camera for approximately 2 seconds. Make sure there is
adequate provision for social distancing in your chosen measurement location.
6)
The camera must have a clear, unobstructed view of the Subject's face, ideally
with the camera at a height around eye level. If necessary, use one of the
mounting accessories available from AMETEK Land (see Section 1.3).
7)
The Blackbody Reference Source requires a mains voltage supply. You will also
need to provide power to the Laptop/PC running the VIRALERT 3 software.
8)
Once the system is installed and operating, use cable ties to secure the
cables to suitable structures. This minimises the risk of the cables becoming
dislodged and reduces the risk of tripping over the cables.
9)
Posters which explain the screening process and provide instructions for your
staff and visitors on how to use the system are available for download from:
www.landviralert.com

3.2 System Connections

NOTE
If you have purchased a mounting accessory i.e. Surface Mount Bracket, Clamp
Mount Bracket, or Tripod from AMETEK Land, then each accessory will come supplied
with its own installation instructions.
04
QUICK START GUIDE
HUMAN BODY TEMPERATURE SCREENING SYSTEM
Camera Connection
Connect the cylindrical connector on the
power and signal cable to the Camera