For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
portable and mobile RF communications equipment and the [KD-5905]
The [KD-5905] is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the [KD-5905] can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and
the [KD-5905] as recommended below, according to the maximum output power of the communications
equipment.
Rated maximum output
power of transmitter
W
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
meters (m) can be determined using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
DN:KD-5905-SMSY01 V1.0
Recommended separation distances between
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d
1
2 .
0.12
0.38
1.2
3.8
12
Table 4
m
80 MHz to 800 MHz
P
d
1
2 .
0.12
0.38
1.2
3.8
12
16 / 16
Operation Guide
800 MHz to 2,5 GHz
P
d
2
3 .
0.23
0.73
2.3
7.3
23
Template version:1
P