Absolute Process Instruments Cecomp Test+ CTP3B1000PSIG Instrucciones
Navegue en línea o descargue pdf Instrucciones para Instrumentos de medida Absolute Process Instruments Cecomp Test+ CTP3B1000PSIG. Absolute Process Instruments Cecomp Test+ CTP3B1000PSIG 4 páginas. Battery powered digital pressure gauges, user configurable
Cecomp
Test + Battery Powered Digital Pressure Gauges, User Configurable
®
Ranges and Resolution
See table below for ranges
Default engineering unit: psi
User configurable for other engineering units
Resolution is fixed for each engineering unit
Accuracy
Accuracy includes linearity, hysteresis, repeatability
Sensor hysteresis:
±0.015% FS, included in accuracy
Sensor repeatability: ±0.01% FS, included in accuracy
Pressure and absolute ranges
±0.1% full scale ±1 least significant digit
Vacuum and vacuum/pressure ranges
±0.25% full scale ±1 least significant digit
Display
3 readings per second nominal display update rate
4 digit LCD, 0.5"H and 5 character 0.25"H alphanumeric
White LED backlight active with button keypress
Batteries, Battery Life, Low Battery Indication
2 AA alkaline included
Approx. 150-1500 hours depending on backlight usage
Low battery symbol on display
Controls and Functions
Three button keypad
Zero/clear/up
On/off
Memory/down
Passcode protected calibration, engineering unit selection, auto
shutoff time, min/max setup
Min/Max Memory
Minimum and/or maximum readings stored in memory
Readings cleared or stored at shutoff
User configurable
Vacuum
Range/Resolution
CTP3B15PSIVAC
0 to 14.70 psig vacuum
Vacuum/Pressure
Range/Resolution
CTP3B±15PSIG
–14.7 to 15.0 psig
CTP3B15V100PSIG
–14.7 to 100.0 psig
CTP3B15V200PSIG
–14.7 to 200.0 psig
Pressure
Range/Resolution
CTP3B3PSIG
0 to 3.000 psig
CTP3B5PSIG
0 to 5.000 psig
CTP3B15PSIG
0 to 15.00 psig
CTP3B30PSIG
0 to 30.00 psig
CTP3B60PSIG
0 to 60.00 psig
CTP3B100PSIG
0 to 100.0 psig
CTP3B200PSIG
0 to 200.0 psig
CTP3B300PSIG
0 to 300.0 psig
CTP3B500PSIG
0 to 500.0 psig
CTP3B1000PSIG
0 to 1000 psig
CTP3B2000PSIG
0 to 2000 psig
CTP3B3000PSIG
0 to 3000 psig
CTP3B5000PSIG
0 to 5000 psig
Absolute Reference
Range/Resolution
CTP3B15PSIA
15.00 to 0 psi absolute
CTP3B30PSIA
30.00 to 0 psi absolute
CTP3B100PSIA
100.0 to 0 psi absolute
Calibration Documentation Options—add to end of model number
CD
Calibration data; 5 test points and date
Accessories—order separately
DPG-OK2, DPG-OK3, DPG-OK6
Pelican
brand high visibility orange heavy
®
duty waterproof cases. Models available for
storing 2, 3, or 6 gauges.
C E CO MP
Div. of
Calibration
Zero button for gauge reference ranges
Non-interactive zero, span, and linearity, ±10% of range
Auto Shutoff
Factory default: 5 minutes
User selectable 1 minute to 8 hours or front button on/off
Weight
Gauge:
9 ounces (approximately)
Shipping:
1 pound (approximately)
Housing Materials
Epoxy powder coated aluminum case with case stiffener,
ABS/ polycarbonate bezel, coated circuit boards for humidity
resistance, front and rear rubber gaskets, polycarbonate label
NEMA 2
Includes rubber boot
Connection and Material
1/4" NPT male fitting
All wetted parts are 316L stainless steel
Overpressure, Burst, Vacuum Service
3000 psig sensor: 5000 psig overpressure
5000 psig sensor: 7500 psig overpressure
All others:
Burst pressure:
Vacuum service:
Environmental Temperatures
Storage:
Operating:
Compensated range: 32 to 158°F
Dimensions
3.67"W x 3.19"H x 2"D with boot, not including fitting
Accuracy
±0.25% FS ±1 LSD
Accuracy
±0.25% FS ±1 LSD
±0.25% FS ±1 LSD
±0.25% FS ±1 LSD
Accuracy
±0.25% FS ±1 LSD
±0.25% FS ±1 LSD
±0.1% FS ±1 LSD
±0.1% FS ±1 LSD
±0.1% FS ±1 LSD
±0.1% FS ±1 LSD
±0.1% FS ±1 LSD
±0.1% FS ±1 LSD
±0.1% FS ±1 LSD
±0.1% FS ±1 LSD
±0.1% FS ±1 LSD
±0.1% FS ±1 LSD
±0.1% FS ±1 LSD
Accuracy
±0.1% FS ±1 LSD
±0.1% FS ±1 LSD
±0.1% FS ±1 LSD
CON14SS
Quick connector to install or remove gauge
without tools. 304 stainless steel, urethane
seal.
A
P
BSOLUTE
ROCESS
2 X pressure range overpressure
4 X sensor pressure rating, or 10,000 psi,
whichever is less
15 psia, ±15 psig, 15 psig, 30 psia,
100 psig, 100 psia, 200 psig sensors
–40 to 203°F (–40 to 95°C)
–4 to 185°F (–20 to 85°C)
(0 to 70°C)
psi, inHg, torr, mmHg, inH2O, ftH2O, oz/in2 cmH2O, g/cm2, kg/cm2, atm, bar, mbar, MPa, kPa
psi, inHg/psi, inHg, torr, mmHg, inH2O, ftH2O, oz/in2, cmH2O, g/cm2, kg/cm2, atm, bar, mbar, MPa, kPa
psi, inHg/psi, inHg, torr, mmHg, inH2O, ftH2O, oz/in2, cmH2O, g/cm2, kg/cm2, atm, bar, MPa, kPa
psi, inHg/psi, inHg, torr inH2O, ftH2O, oz/in2, kg/cm2, atm, bar, MPa, kPa
psi, inHg, torr, mmHg, inH2O, ftH2O, oz/in2, mmH2O, cmH2O, g/cm2, kg/cm2, atm, bar, mbar, kPa
psi, inHg, torr, mmHg, inH2O, ftH2O, oz/in2, mmH2O, cmH2O, g/cm2, kg/cm2, atm, bar, mbar, kPa
psi, inHg, torr, mmHg, inH2O, ftH2O, oz/in2, cmH2O, g/cm2, kg/cm2, atm, bar, mbar, MPa, kPa
psi, inHg, torr, mmHg, inH2O, ftH2O, oz/in2, cmH2O, g/cm2, kg/cm2, atm, bar, mbar, MPa, kPa
psi, inHg, torr, mmHg, inH2O, ftH2O, oz/in2, cmH2O, g/cm2, kg/cm2, atm, bar, mbar, MPa, kPa
psi, inHg, torr, mmHg, inH2O, ftH2O, oz/in2, cmH2O, g/cm2, kg/cm2, atm, bar, mbar, MPa, kPa
psi, inHg, inH2O, ftH2O, oz/in2, kg/cm2, atm, bar, MPa, kPa
psi, inHg, ftH2O, oz/in2, kg/cm2, atm, bar, MPa, kPa
psi, inHg, ftH2O, kg/cm2, atm, bar, MPa, kPa
psi, inHg, ftH2O, kg/cm2, atm, bar, MPa, kPa
psi, inHg, ftH2O, kg/cm2, atm, bar, MPa
psi, inHg, ftH2O, kg/cm2, atm, bar, Mpa
psi, kg/cm2, atm, bar, Mpa
psi, inHg, torr, mmHg, inH2O, ftH2O, oz/in2, cmH2O, g/cm2, kg/cm2, atm, bar, mbar, MPa, kPa
psi, inHg, torr, mmHg, inH2O, ftH2O, oz/in2, cmH2O, g/cm2, kg/cm2, atm, bar, mbar, MPa, kPa
psi, inHg, torr, mmHg, inH2O, ftH2O, oz/in2, cmH2O, g/cm2, kg/cm2, atm, bar, mbar, MPa, kPa
NC
NIST traceability documentation, 5 points and date
1220 American Way Libertyville, IL 60048
I
NSTRUMENTS
cecomp.com/plus
O ±0.1% Test Gauge Accuracy in Most Ranges
O Ruggedized Design
O Red Rubber Boot Included
O 316L Stainless Steel Wetted Parts
O Selectable Units and Auto Shutoff Times
O White LED Display Backlight
O Min/Max Memory
Pb
Lead
Free
Selectable Units
Selectable Units
Selectable Units
Selectable Units
SCR14SS
Filter screen fitting keeps debris out of gauge
sensor. For food vacuum packaging applica-
tions. 303SS body, 100 micron 304SS screen.
cecomp.com
800-942-0315
CTP3B
H H H H H H
H H H H H
H H H H H H
H H H H H
H H H H H H
H H H H H
H H H H H H
H H H H H
H H H H H H
Made in USA
© 10-19