Apogee Instruments SU-300 UV-I Manual del usuario - Página 7
Navegue en línea o descargue pdf Manual del usuario para Accesorios Apogee Instruments SU-300 UV-I. Apogee Instruments SU-300 UV-I 19 páginas. Ultraviolet index & ultraviolet-b sensor
SPECIFICATIONS
Power Supply
Output (sensitivity)
Calibration Factor (reciprocal of
sensitivity)
Calibration Uncertainty
Output Range
Measurement Range
Measurement Repeatability
Long-term Drift
Non-linearity
Response Time
Field of View
Spectral Range
Directional (Cosine) Response
Temperature Response
Operating Environment
Dimension
Mass (5 m of cable)
Cable
Calibration Traceability
Apogee UV-I/UV-B series sensors are calibrated through side-by-side comparison to the mean of four transfer
standard UV-I/UV-B sensors under sunlight. The transfer standard UV-I/UV-B sensors are calibrated through side-
by-side comparison to an Apogee model PS-300 spectroradiometer under sunlight (clear sky conditions) in Logan,
Utah. The PS-300 is calibrated with a quartz halogen lamp traceable to the National Institute of Standards and
Technology (NIST).
270 to 315 nm (wavelengths where response is greater than 50 % of maximum; see Spectral
-30 to 85 C; 0 to 100 % relative humidity
5 m of shielded, twisted-pair wire; TPR jacket (high water resistance, high UV stability,
flexibility in cold conditions); pigtail lead wires; stainless steel (316), M8 connector located
SU-300-SS
Self-Powered
0.1 mV per UV Index unit
10 UV Index units per mV
± 10 %
0 to 2 mV
0 to 20 UV Index
Less than 0.5 %
Less than 2 % per year
Less than 1 %
Less than 1 ms
180°
Response graph below)
± 2 % at 45°, ± 5 % at 75°
Less than 0.1 % per C
30.5 mm diameter, 37 mm height
140 g
25 cm from sensor head