9. Error Messages
If an error occurs during the measurement, the
measurement is interrupted and an error message,
e.g. «ERR 3», is displayed.
Error
Designa-
tion
"ERR 1" Signal
too weak
"ERR 2"
Error
signal
18
No pres-
"ERR 3"
sure in the
17
cuff
11
Potential cause
and remedy
The pulse signals on
the cuff are too weak.
Re-position the cuff and
repeat the measurement.*
During the measurement,
error signals were de-
tected by the cuff, caused
for instance by movement
or muscle tension.Repeat
the measurement, keeping
your arm still.
An adequate pressure
cannot be generated in
the cuff. A leak may have
occurred. Check that the
cuff is correctly connected
and is not too loose.
Replace the batteries if
necessary. Repeat the
measurement.
Error
Designa-
tion
Abnormal
"ERR 5"
result
Pulse or
cuff pres-
"HI"
sure too
high
Pulse too
"LO"
low
* Please consult your doctor, if this or any other
problem occurs repeatedly.
If you think the results are unusual, please
read through the information in «Section 1.»
carefully.
Potential cause
and remedy
The measuring signals
are inaccurate and no
result can therefore be
displayed. Read through
the checklist for perform-
ing reliable measurements
and then repeat the
measurement.*
The pressure in the cuff
is too high (over 300
mmHg) OR the pulse is
too high (over 200 beats
per minute). Relax for 5
minutes and repeat the
measurement.*
The pulse is too low (less
than 40 beats per minute).
Repeat the measure-
ment.*