ADE M400020 Series Manual de instrucciones - Página 12

Navegue en línea o descargue pdf Manual de instrucciones para Scales ADE M400020 Series. ADE M400020 Series 16 páginas. Class iii approved scales electronic chair scales

ADE M400020 Series Manual de instrucciones

Recalibration

Have the calibration carried out by an authorised company or authorised person according to national legisla-
tive regulations. The CE symbol shows the notified body (0122) and the year of initial calibration. Recalibration
must be carried out
-
if the calibration mark has been damaged
-
after calibrated scales have been successfully repaired
-
or after expiry of a period for recalibration as stipulated by national, legal regulations.
Power supply
Mains Adapter:
Mains voltage
Frequency
Power Consumption
Main power circuit
Input Voltage
Max. Current (in-charger)
Battery Requirement:
Battery Operation:
Charger
Saving (350ohm, Backlight-ON 30Sec,
Shutdown 5Min)
Serial Communications
EDP Port
Printer Port
Both Ports
Operator Interface
Display
LCD annunciator
Keyboard
Environmental
Operating Temperature
Storage Temperature
Humidity
Altitude
Housing
Dimensions
Weight
Material
12

Technical specifications

115 or 230VAC
50 or 60HZ
11W in-charger, 2W off-charger
5.0VDC
1000mA
Rechargeable Lithium Ion Cylindrical (18650x2) Cell 3.7V 4000mAH
7~10 Hours
200Hours
Full duplex RS-232
38400,19200,9600, 4800, 2400, 1200 bps
Output-only RS-232 or active 20 mA current loop,
9600, 4800, 2400, 1200 bps
7 or 8 data bits; even, odd, or no parity
6-digit LCD display. 7-segment, 0.815 in (21.2 mm) digits
net, center of zero, standstill, kg/lb/g/oz/ units,
Battery, Set points, Hold, HI/OK/LO
24-key flat membrane panel
–10 to +40°C (legal);
–25 to +70°C
0–95% relative humidity
2000 m (6500 ft) maximum
18.6 cm x 9.5 cm x 10 cm
0.4 kg (without battery)
ABS plastic, white
M400020-210322-Rev007-UM-e