Printer connections (TxD 20 mA, either a printer or remote display can be used)
To Personal Computer (RS232C, ADE- MAK Stream Output)
To Printer / Remote Display (RS232C, ADE- MAK Print Function/Stream Output)
To Printer / Remote Display (+20mA, ADE- MAK Print Function/Stream Output)

Housing and connections

The mains connection, serial connection and weighing cell connection points are located on the back
of the ADE-MAK housing. The set-up switch to start configuration mode is recessed underneath the
housing. It is protected by a cover plate and secured by a cylinder head screw (not shown in Fig. 2-1).
The PIN allocation of the connecting sockets is shown in Fig. 2-2.
Do not press the set-up switch as otherwise calibration will be lost (reserved for use by
service personnel only)!
Manipulations made to test and configuration modes will result in the forfeiture of any
warranty claims!
A PC, printer or digital inputs and outputs can be connected through the SUB-D connecting socket.