Great Plains Industries FM-300H Manual del usuario - Página 10

Navegue en línea o descargue pdf Manual del usuario para Instrumentos de medida Great Plains Industries FM-300H. Great Plains Industries FM-300H 16 páginas. Electronic disc meter

Great Plains Industries FM-300H Manual del usuario
Cleaning
During use, the meter should be kept
free of liquid to ensure drying does not
occur inside the meter. If drying should
occur, the nutator will stick or drag,
affecting accuracy. If this occurs, clean-
ing is required.
Cleaning the Nutator
1. Loosen the screws on the back of
the meter about 1/4 inch (0.6 cm).
(Figure 4)
2. Tap the screws with a hammer to
loosen the coverplate from the hous-
ing. (Figure 5) Remove the screws
and washers.
3. Gently pry the coverplate from the
housing. Take care not to damage
the coverplate, housing, or O-ring.
(Figure 6)
10
4. Remove the signal generator from
the pin on the nutator disc.
Figure 4
(Figure 7)
5. Gently remove the metal clips hold-
ing the nutator in place. (Figure 8)
Take care not to damage the O-ring
Figure 5
on the nutator inlet.
6. Remove the nutator.
7. Using water or a solvent-based fluid
and a brush, carefully remove resi-
due from the nutator disc, nutator
halves, signal generator, and inside
of housing. Allow all parts to dry.
Figure 6
Figure 7
Figure 8