iHealth PT1 Manual del usuario - Página 9

Navegue en línea o descargue pdf Manual del usuario para Termómetro iHealth PT1. iHealth PT1 10 páginas. Clinical digital thermometer

iHealth PT1 Manual del usuario

Electronic Compatibility Information

Table 1 - Emission limits per environment
Phenomenon
Conducted and
radiated RF
emissions
Harmonic
distortion
Voltage
uctuations
and icker
Table 2 - Enclosure Port
Phenomenon
Electrostatic
Discharge
Radiated RF EM eld
Proximity elds from
RF wireless
communications
equipment
Rated power
frequency
magnetic elds
Table 3 - Proximity elds from RF wireless communications
equipment
Test frequency
(MHz)
385
450
710
745
780
810
870
930
1720
1845
1970
2450
5240
5500
5785
Compliance
CISPR 11
Group 1, Class B
IEC 61000-3-2
NA
IEC 61000-3-3
NA
Basic EMC
standard
IEC 61000-4-2
IEC 61000-4-3
IEC 61000-4-3
IEC 61000-4-8
Band (MHz)
380-390
Pulse modulation 18Hz, 27V/m
430-470
FM, ±5kHz deviation, 1kHz sine, 28V/m
704-787
Pulse modulation 217Hz, 9V/m
800-960
Pulse modulation 18Hz, 28V/m
1700-1990
Pulse modulation 217Hz, 28V/m
2400-2570
Pulse modulation 217Hz, 28V/m
5100-5800
Pulse modulation 217Hz, 9V/m
8
Electromagnetic
environment
The device is intended
to be used in home
healthcare environment.
The device is
powered by battery.
The device is
powered by battery.
Immunity test levels
Home Healthcare
Environment
±8 kV contact
±2kV, ±4kV, ±8kV, ±15kV air
10V/m
80MHz-2.7GHz
80% AM at 1kHz
Refer to table 3
30A/m
50Hz or 60Hz
Immunity test levels
Professional healthcare facility
environment