Aqua Lab PRE Manual de inicio rápido - Página 6

Navegue en línea o descargue pdf Manual de inicio rápido para Instrumentos de medida Aqua Lab PRE. Aqua Lab PRE 17 páginas. Water activity meter

Aqua Lab PRE Manual de inicio rápido
Daily Verification- Step Two

Determine Second Standard

5
Need Vials? Go to https://store.aqualab.com
1
If your primary verification
standard read within
+/- 0.01 of the given value
for that standard, identify
the range of water activity
you plan to test and whether
it is above or below your
primary standard.
2
Choose a second verification
standard that borders the range
of water activity you plan to
test. For example, if you plan to
test for water activity readings
ranging between 0.71 and
0.62 you should use the 0.760
as your primary standard and
the 0.500 as your secondary
standard.
3
Conduct a two-point
performance verification daily.
If you need guidance on which
two standards to use, please
contact us:
Phone: 509-332-5601