Olympus BHS Manual de instrucciones - Página 12
Navegue en línea o descargue pdf Manual de instrucciones para Microscopio Olympus BHS. Olympus BHS 22 páginas. System microscope
También para Olympus BHS: Manual de instrucciones (22 páginas)
©
An
Olympus
objective
engraved
'160/0.17"
requires a cover glass of 0.17mm
thickness.
If the numerical
aperture of the objective is 0.7 or higher (except immersion objectives) and
no correction collar is provided, the resolving power deteriorates to a great extent if cover
glass thickness deviates from the above
listed value.
NOTE:
In some countries a 0.17mm
cover glass corresponds to a designation of #1%.
@
A cover glass (0.4mm
thick)
for blood counting, etc. can be used with Olympus objectives
of the working distance longer than 0.4mm.
(See VI.
OPTICAL
DATA at page 16.)
[Specimen Slide
®@
Specimen
slides 0.8mm
to 1.2mm
thick are
recommended
for Olympus objectives.
3) Bring the portion of the specimen for obser-
vation
into the
light
path
by
means
of the
low drive control knobs. (Fig. 9)
*
Tighten the stage clamping screw © in the
microscope front.
[Stage
© The specimen holder can accommodate two
standard specimen slides simultaneously.
@ The specimen holder is removable to obtain
a large
unobstructed
stage
surface
to hold
specimens up to 55mm x 85mm.
@
To rotate the stage loosen the stage clamping
screw ©
and holding this screw, rotate the
stage into the desired direciton.
(Fig. 9)
O Stage clips @®, for use with immersion objec-
tives.
(Fig. 10)
A pair of stage clips are optionally
available
to hold the specimen on the stage, eliminat-
ing a specimen drag caused by immersion oil
between
slide and
stage surface.
The clips
can
be inserted
into the holes provided
on
the specimen holder.
Fig. 10