Accuris W3200-1200 Manual de instrucciones - Página 11
Navegue en línea o descargue pdf Manual de instrucciones para Scales Accuris W3200-1200. Accuris W3200-1200 16 páginas. Precision balance
También para Accuris W3200-1200: Manual de instrucciones (16 páginas)
COUNT/ENTER to confirm. "0 SEC" indicates continuous print.
5.Print COUNT/ENTER key to return to weighing mode.
6.After entering print setup menu, press TARE key repeatedly to choose line feed setup
"LinEFd".
7.Press TARE key repeatedly to see the predetermined line feeds (1-18). Choose the
proper line feeds, press the COUNT/ENTER key to return to weighing mode.
Baud Rate Setup
1.Press the MENU key.
2.Press the TARE key, until "bAud" is displayed.
3.Press the COUNT/ENTER key, the balance reads "300".
4.Press the TARE key repeatedly, the balance displays the other buad rates. Choose the
proper rate and press COUNT/ENTER to confirm. The balance reads "ParITy" and
begins parity check.
5.The balance reads "nOnE" (no check) for the first.
6.Press the TARE key, it reads the check type, "Odd" indicates odd check and "EvEn"
indicates even check.
7.Choose the proper check type and press COUNT/ENTER to confirm, the balance
now returns to weighing mode.
Enabling Units of Measure
The units function can be programmed to turn certain weighing units on or off. To enable
or disable certain units of measure, perform the following procedure.
1.Press MENU key, the display reads "PrInT".
2.Press TARE key repeatedly until it reads "unIT".
3.Press COUNT/ENTER key, the display reads "g yes" which means g available for
use. Press COUNT/ENTER to confirm. To disable g as the unit, press Tare key, the
display reads "g no", press COUNT/ENTER key to confirm.
4.Follow the steps above to enable or disable Oz, Ct or dwt as the unit of measure.
Restoring the Factory Default Setup
This step resets the factory defaults.
1. Press the Menu key, the display reads "PrInT".
2. Press Tare key repeatedly until it reads "InITIA".
3. Press Count/Enter key to confirm, the display reads "BUSY" and then returns to
weighing mode. Factory default setup is finished.
Backlight Setup
1.Press MENU key, the display reads "PrInT".
2.Press TARE key repeatedly until it reads "bLgHT", press COUNT/ENTER to
confirm, the display reads "1 nIn" which indicates the backlight will be off in 1
minute.
3.Press TARE key repeatedly to choose the desired time, 1、2、3、5、10、15、30、
60 (min. are available to choose).
4.Press COUNT/ENTER key to confirm and the balance returns to weighing mode.
11