AccuVein AV500 Manual del usuario - Página 13

Navegue en línea o descargue pdf Manual del usuario para Equipamiento médico AccuVein AV500. AccuVein AV500 17 páginas. A health professional's guide for use and operation

AccuVein AV500 Manual del usuario
Recommended separation distances between portable and mobile RF communications equipment and the AV500
The AV500 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the AV500 can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the AV500 as recommended
below, according to the maximum output power of the communications equipment.
Rated Maximum Output power of
Separation distance according to frequency of transmitter, m
transmitter, W
150kHz to 80MHz
D = 1.2P
0.01
0.12
0.1
0.38
1
1.2
10
3.8
100
12
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.
Guidance and Manufacturer's Declaration—Electromagnetic Emissions
The AV500 is intended for use in the electromagnetic environment specified below. The customer or the user of the AV500 should assure that it is used in such an environment.
Emissions Test
Compliance
RF Emissions
Group 1
CISPR 11
RF Emissions
Class A
CISPR 11
Harmonic Emissions
Complies
IEC 6100-3-2
Voltage Fluctuations
Complies
IEC 6100-3-3
80MHz to 800MHz
800MHz to 2.5GHz
0.12
0.12
0.38
0.38
1.2
1.2
3.8
3.8
12
12
Electromagnetic Environment Guidance
The AV500 uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any
interference in nearby electronics.
The AV500 is suitable for use in all establishments other than domestic and those directly connected to the public low-voltage
power supply network that supplies buildings used for domestic purposes.
EMI/EMC
Running head
13