StarClose SE
Vascular Closure System
"CLICK" 2 - Locator Wing Deployment
Hold the Stabilizer with the left hand. Stabilize the body of the Clip Applier at the angle of the tissue tract.
With the right index and middle fingers on the proximal posts, right thumb on the Plunger, retract the Clip Applier out of the tissue 3 to
4 centimeters.
Using the right thumb, depress the Plunger firmly to deploy the Locator Wings, which advances the Thumb Advancer about
3 centimeters. The initial splitting of the Sheath will be visible above the skin level.
A "CLICK" should be heard to confirm Locator Wing deployment. Check that the number "2" is completely visible in the number window.
©2011 Abbott. All rights reserved. ANZ 080091 10/11