APPA A17 Carte d'instruction - Page 3
Parcourez en ligne ou téléchargez le pdf Carte d'instruction pour {nom_de_la_catégorie} APPA A17. APPA A17 3 pages. Leakage current clamp meter
Auto Hold
Battery Replacement
Maintenance
Do not attempt to repair this Tester.
It contains no user-serviceable parts.
Repair or serving should only be
performed by qualified personnel.
Cleaning
Periodically wipe the case with a dry
cloth and detergent do not use
abrasives or solvents.
Specifications
1-1 General Specifications
LCD display digits :
6000 digit large scale LCD readout.
Measuring rate : 5 times / sec.
Overrange display :
Shows the real value for "100A" and
shows "OL" for other ranges.
Automatic power off time :
Approximately 20 minutes after power on
without operations.
Low battery indicator :
is displayed.
Replace the battery when the indicator
appears in the display.
Power requirement : 1.5Vx2 batteries.
Battery type and life :
ALKALINE 60 hours.
1-2 Environmental Conditions
Indoor Use.
Pollution degree : 2
Calibration : One year calibration cycle.
Operating temperature :
0°C ~ 30°C (≦80% RH)
30°C ~ 40°C (≦75% RH)
40°C ~ 50°C (≦45%RH)
Storage temperature : -20 to +60°C,
0 to 80% RH (batteries not fitted).
Operating altitude : 2000m (6562 ft)
Conductor Size : 40mm diameter.
Shock vibration :
No shock, No vibration to the jaw tips.
Temperature coefficient :
0.2 x (Specified accuracy) / °C, < 18°C,
> 28°C .
EMC : EN 61326-1
Overvoltage category :
IEC 61010-1 300V CAT.Ⅲ
CAT
Application field
Ⅰ
The circuits not connected to mains.
The circuits directly connected to
Ⅱ
Low-voltage installation.
Ⅲ
The building installation.
The source of the Low-voltage
Ⅳ
installation.
1-3 Electrical Specifications
Accuracy is ±(% reading + number
of digits) at 23°C ± 5°C < 80%RH.
The specifications of 50 ~ 100A is not
Specified.
Current
Range
Resolution
Accuracy
6mA *
0.001mA
0 ~ 50A
60mA
0.01mA
1% + 3D (50 ~ 60Hz)
600mA
0.1mA
50 ~ 60A
5% + 6D (50 ~ 60Hz)
6A
0.001A
60 ~ 100A
60A
0.01A
8% + 6D (50 ~ 60Hz)
100A
0.1A
* There is a little rolling within ±3 digits.
Frequency Response (61~400Hz) :
2%+6D(0~30A), 30%+6D(30~100A)
Influence of external magnetic field of
adjacent wires : 0.04% typical value.
Low Pass Filter
Range
Resolution
Accuracy
6mA *
0.001mA
0 ~ 50A
60mA
0.01mA
2% + 6D (50 ~ 60Hz)
600mA
0.1mA
50 ~ 60A
6% + 8D (50 ~ 60Hz)
6A
0.001A
60 ~ 100A
60A
0.01A
9% + 8D (50 ~ 60Hz)
100A
0.1A
* There is a little rolling within ±3 digits.
Attenuation characteristic:
approx. –24dB/oct
Cut-off frequency : 100Hz
Limited Warranty
This Meter is warranted to the original
purchaser against defects in material
and workmanship for 2 years from the
date of purchase. During this warranty
period, manufacturer will, at its option,
replace or repair the defective unit,
subject to verification of the defect or
malfunction. This warranty does not
cover disposable batteries, or damage
from abuse, neglect, accident, unauthor-
ized repair, alteration, contamination, or
abnormal conditions of operation or
handling.
Any implied warranties arising out of
the sale of this product, including but
not limited to implied warranties of
merchantability and fitness for a
particular purpose, are limited to the
above. The manufacturer shall not be
liable for loss of use of the instrument
or other incidental or consequential
damages, expenses, or economic loss,
or for any claim or claims for such
damage, expense or economic loss.
Some states or countries laws vary, so
the above limitations or exclusions may
not apply to you.
APPA TECHNOLOGY CORP.
9F, 119-1 Pao-Zong Rd.,
Shin-Tien, Taipei, 23115, Taiwan.
P.O.Box. 12-24 Shin-Tien, Taiwan.
Tel: +886-2-29178820
Fax: +886-2-29170848
E-mail: [email protected]
http: //www.appatech.com