ADE KWE15-IP65 Manuel d'utilisation - Page 5

Parcourez en ligne ou téléchargez le pdf Manuel d'utilisation pour {nom_de_la_catégorie} ADE KWE15-IP65. ADE KWE15-IP65 16 pages. Compact scale in stainless steel housing

ADE KWE15-IP65 Manuel d'utilisation
2.6

Switch on

Press [ON]. The whole display illuminates with a serial of steps and ends with the screen
display „0". After the [ZERO] indicator is lighting, the scale is ready for weighing.
2.6.1. Switch-off
Scale can be turned off forcibly by pressing [OFF].
2.6.2. Auto-off / Standby (only working with rechargeable batteries)
The scale has an automatic shut-off function, which can be activated or deactivated in
parameter setting.
Setting could be automatic shut-off after 10 minutes or deactivated shut-off. In this case
depending on the setting the backlight is always on or shut-off after 40 seconds showing "0".
If you place something on the platform, the scale goes back in weighing modus.
2.6.3. Low voltage auto-off
If the voltage of the rechargeable batteries is less than 5.6V, the scales turn off automatically.
2.7

Zero Operation

When scale is on, [ON] is the key to zero. It is only available within 4% of the maximum
capacity; otherwise it will immediately display "HHHHHH".
2.8

Displaying the voltage

When scale is on, it will display the voltage when pressing [ON] key for 3 seconds. Now if press
[ON] again, the scale will return to weighing mode.
2.9

Tare Operation

Place the container on the scale until the display indicates the weight number steadily.
Press [TARE], then the display indicates "0" and the TARE indicator is lighting.
Place the weighing objects into the container, and then scale shows the Net weight.
Move the container and the weighing objects off, and then scale will show the minus weight of
the container. Then press [TARE] again, the tare weight can be cleared, that the TARE indicate
out and the scale is set to zero.
If the display is not stable, no arrow is shown.
KWE-190107-Rev008-UM-en
4