iHealth PT2L Manuel de l'utilisateur - Page 11
Parcourez en ligne ou téléchargez le pdf Manuel de l'utilisateur pour {nom_de_la_catégorie} iHealth PT2L. iHealth PT2L 15 pages. Infrared digital no-touch thermometer
Care and Cleaning
1.
The probe (sensor) is the most intricate part of the thermometer and should be kept
clean and intact to acquire accurate readings.
Use the following method to clean the probe:
1)
Gently swab the surface of the probe using a cotton bud soaked in > 95%
medical
2)
Allow at least 1 minute for the probe to fully dry.
2.
If the probe (sensor) is broken, please contact customer services.
3.
Use a piece of soft, dry cloth to clean the display screen and external surface of the
thermometer. If the thermometer is very dirty, the cloth can be moistened with some
medical alcohol to clean the device.
4.
The product is not waterproof. Do not clean the device with detergents. Do not soak
the thermometer in water or other liquids.
Maintenance
1.
This company has not authorized any agency or individual to carry out product
repairs or maintenance. Do not attempt to disassemble or modify the thermometer if
you suspect functional issues with the device.
2.
The IR thermometer is an extremely precise instrument. Any improper maintenance,
disassembly, or modification may lead to inaccuracies of the product measurements.
3.
Please check the device if damaged after it is dropped. If unsure, please contact
customer services for having the device checked.
4.
The thermometer is initially calibrated at the time of manufacture. If this thermometer
is used according to the use instruction, periodic re-adjustment is not required. If any
time your question the accuracy of measurement during the warranty period, please
contact customer services.
5.
No component can be maintained by user in the monitor. circuit diagrams,
component part lists, descriptions, calibration instructions, or other information which
will assist the user's appropriately qualified technical personnel to repair those parts
of equipment which are designated repairable can be supplied.
6.
The method for verifying the clinical accuracy can be requested, please contact
customer services.
7.
The device must not be stored or used at an excessively high or low temperature or
humidity (see technical data), in sunlight, in association with an electrical current or
in dusty locations. Avoid dropping or subjecting the product to external forces.
Otherwise inaccuracies can occur.
8.
Do not directly touch the probe with your fingers or blow on it. Measurements taken
using a damaged or dirty IR probe may be inaccurate.
9.
The monitor requires 4 hours to warm from the minimum storage temperature between
uses until the monitor is ready for its INTENDED USE when the ambient temperature is
20 °C (68 °F ).
10. The monitor requires 4 hours to cool from the maximum storage temperature between
uses until the monitor is ready for its INTENDED USE when the ambient temperature is
alcohol.
9