FM 10-67-1 APPENDIX E
500 mL sample, the scale reads directly in PPM of free water in the fuel.
If your reading is 10 PPM or below, the test is finished and the fuel may
be used for Army or Air Force aircraft. Readings of 5.0 PPM or below
are required for use in Navy and Marine Corps aircraft.
If the reading is off the scale (on the high side), repeat sampling and
testing procedures using a sample volume below 500 mL but no lower
than 100 mL. Record the reading from the point where the lever crosses
the scale, when the meter pointer points to zero, and perform the
following calculation:
Free water, ppm = (meter reading, ppm) (500)/(sample volume, mL)
If the retest shows more than the 10 PPM of water, submit a sample to the
supporting laboratory for inspection and disposition instructions and take the
fuel and refueler or fuel system out of service.
Keep the kit and its components clean. Store the components in the kit and
replace components when necessary. Proper maintenance is essential. Follow
cleaning and maintenance instructions in TM 105-66430-22118-13&P0.
http://www.globalsecurity.org/military/library/policy/army/fm/10-67-1/APPE.HTML

CLEANING AND MAINTENANCE