Measurement Screen

Reading in Process- sunburst icon above
Reading in process
Reading is still in process if you see
the sunburst
5
a
w.
icon.
Final Reading– no sunburst icon above
Reading complete
Reading is complete when your Pawkit 'beeps'
and the sunburst icon disappears.
If the standard measures between 0.74
and 0.78, move on to the second verification
standard. If not, clean the instrument and retry.
*Accuracy of the Pawkit ± 0.02
a
w.