7.
Remove and retain the tie bar fixing bolts from the top flange and remove the top flange.
Figure 11 Disassembly of the Column Tube Unit
8.
Remove tie bars to allow the tube to be lifted out of the bottom tube flange. If tube is difficult to remove from
flange remove the lower buffer seal.
9.
Discard the tube and buffer seals.
®
Evolve
350mm Column: Instructions for Replacing the Wetted Flowpath using the Column Refresh Kit
Figure 10 Removal of tube unit
from Column Base
11