Calibration mode: Only for service
Mode E:
Mode display
The letter 'E' is shown in the mode display.
The small + large displays show the count value of the internal A/D converter. This
mode is only used for service work.
IMPORTANT:
If you enter calibration mode accidentally, hold the minus key pressed until the
software returns to the normal operating mode (Mode 0).

Maintenance, disposal

Please observe the following instructions :
Spent batteries are hazardous waste and must be disposed as prescribed by law.
Do not use carbon-containing chemicals or fuel, alcohol or similar fluids for cleaning.
• • • •
Damage due to non-compliance with the above instructions is not covered by
guarantee.

Troubleshooting and remedy

Problem
The device does not display
temperature but only meaningless
characters
Not all modes can be entered
When the device is switched on it does
not go to temperature display mode
(MD3060)
Copyright BECKMANN+EGLE GmbH, All rights reserved.
E
10,000 digit of A/D value
Page 10
Small display
Possible remedy
You are probably in calibration mode. Press the
minus key to return to normal operating mode.
An interval measurement has not been completed
and is still active; see description of mode 3
No sensor connected or sensor defective
Large display
1,000 digit of A/D value