Gamry Instruments Universal Dummy Cell 3 Panduan Operator - Halaman 8

Jelajahi secara online atau unduh pdf Panduan Operator untuk Peralatan Uji Gamry Instruments Universal Dummy Cell 3. Gamry Instruments Universal Dummy Cell 3 8 halaman. Circuit board used for calibration and troubleshooting of a gamry potentiostat.

--DC Corrosion--
A Tafel fit of a typical scan (both with and without iR compensation) can be seen below. Both the anodic and
cathodic Tafel constants were calculated to be between 60 and 70 mV per decade. The calculated corrosion
rate is quite small (about 10
mm/year) because of the very small I
(about 10
Amps/cm
).
-3
-10
2
corr
Note that the Schottky diodes used to produce the linear log current versus potential behavior have a fairly high
variability. Both the current at zero volts and the slope of the log current versus potential plot vary significantly
diode to diode.
For this reason, you should use the results obtained with the DC Corrosion Test cell with some caution. The
shape of the Tafel curve is important, but the values obtained in a Tafel fit are not.
4