Aliyiqi ADT-30 Manual - Halaman 8

Jelajahi secara online atau unduh pdf Manual untuk Peralatan Uji Aliyiqi ADT-30. Aliyiqi ADT-30 12 halaman. Horizontal torsion spring testing machine

Description of setting parameters: (Take the maximum torque value as an
example)
Use the up and down keys to shift and modify the data. At this time, press the set
key to confirm and save the set data, (if you press the zero key to return, the data
will not be modified). The set data cannot exceed the maximum load value.
(1) Maximum torque value: The user sets the upper limit value of the torque,
which can be freely set as required. When the upper limit value is reached, the
number above the parameter name \"torque upper limit value \" flashes and the
buzzer alarms. The upper limit value is not higher than full scale.
(2) Minimum torque value: The user sets the lower limit value of torque, which
can be freely set as required. When the lower limit value is reached, the number
above the parameter name \"torque lower limit value \" flashes. The lower limit
value shall not be higher than the set upper limit value.
(3), minimum storage value setting: the user sets the minimum storage value
according to storage needs, data less than this value will not be stored.
8. Other function description
1. Description of printing function: print data through micro printer.
2. Communication: communicate with the host computer via USB. The
communication protocol adopts MODBUS-RTU protocol.
3. Save data: save the data during measurement through the save key.
4. View stored data: View the stored data through the view button.
5. Alarm description: when the measured value exceeds the upper limit, the
upper limit indicator will be displayed, and the buzzer will also give an alarm;
when the measured value is lower than the lower limit, the lower limit indicator
will be displayed. When the measured value exceeds 120% of the maximum
load, the sensor may be damaged. When the \"Overload \" warning prompt