Chapter 5. Cleaning & Maintain
-Device
(1)
Recommend that pour the liquid in the container before met the full line.
(2)
Disinfected with diluted medical alcohol to clean the device surface after use.
(3)
Disinfected with diluted medical alcohol to clean the accessories, such as suction container, connection
tube, and suction tube...etc.
(4)
Charge the device for 3hrs at least once per month.
(5)
The reusable collection canister is single-patient use and should be replaced every 1 to 2 months.
- Bacteria Filter
(1)
The filter can't be cleaned.
(2)
If the filter gets wet, or becomes contaminated or clogged, it must be replaced.
Chapter 6. Troubleshooting
Problem
Pump runs,
1. Tubing isn't connected properly.
but no vacuum
2. Tubing breaks or leaks.
3. Container do not assembly properly.
4. Container breaks or leaks.
Unit does not
1. Switch broken.
working when
2. PCB board broken.
power is opened
Vacuum
1. Tubing isn't connected properly.
delivered, but
2. Gauge broken
gauge did not
display the
values
Battery will not
1. Charging IC broken
hold a charge.
2. Battery life expires
Low vacuum
1. Tubing isn't connected properly.
2. Pump broken.
3. Suction container full.
Abnormal sound
1. Pump broken
when the device
2. Waste liquid or water
working
Possible Cause
countercurrent to the machine
Solutions
1. Assembly the tube properly.
2. Contact your equipment provider for repair
Contact your equipment provider for repair
Contact your equipment provider for repair
1. Ensure that unit is equipped with battery.
2. Verify that charge light turns on.
3. Check electrical connections during
charging.
4. Contact your equipment provider for repair
1. Assembly the tube properly.
2. Contact your
equipment provider for repair
Contact your equipment provider for repair