Aqua Lab Pawkit Panduan Memulai Cepat - Halaman 8

Jelajahi secara online atau unduh pdf Panduan Memulai Cepat untuk Alat Ukur Aqua Lab Pawkit. Aqua Lab Pawkit 12 halaman. Water activity meter

Aqua Lab Pawkit Panduan Memulai Cepat
Daily Verification-Step Two Continued

Secondary Standard

4
Locate a vial of the Secondary Verification Standard
(0.250, or 0.920) that came with your
instrument or that you purchased separately.
5
Twist off vial cap and pour contents into a new
sample cup.
Take a reading of the second standard by placing the
7
Pawkit over the cup and pressing the left I button.
Error!
If the instrument doesn't measure within specifications,
Standard not
clean the chamber and retry. See Chapter 5 in your
measuring in
Operator' s Manual if the issue persists or contact
specification.
AquaLab Customer Support.
7