AeroClave mRDS Manuale dell'operatore - Pagina 3

Sfoglia online o scarica il pdf Manuale dell'operatore per Apparecchiature per la pulizia AeroClave mRDS. AeroClave mRDS 16. Room decontamination system

INTRODUCTION

1.0

Purpose

1.1
The purpose of this manual, primarily, is to establish guidelines for the safe operation of the mRDS
system. Secondarily, it establishes a method for proper hand-application using the optional APA.

Terms/Definitions

1.2
mRDS:
APA:
Injection Phase:
Dwell Phase:
Aeration Phase:

Equipment and Media

1.3
mRDS Unit
4 Liter AeroClave Solution Tank
AeroClave Approved Solution
APA
Touchscreen Controller
Released January 8, 2016
Use or disclosure of data contained on this sheet is subject to the restriction on the cover page of this User's Manual.
modular Room Decontamination System
AeroClave Portable Applicator
In this phase, the solution is aerosolized and applied evenly to the
treatment area. A typical injection phase can last anywhere from 5-30
minutes, depending on the size of the treatment area.
Once an injection phase has completed, the treatment area is allowed to sit
for a minimum of 10 minutes in a dwell period. The dwell period gives the
aerosolized solution an opportunity to evenly distribute throughout the
treatment area, ensuring complete coverage on all surfaces.
The aeration phase is the final stage when the treatment area is returned to
its normal habitable state. Do not re-enter the area until the air is clear
(minimum 20 minutes).
AeroClave mRDS Operator Manual
Page 3