Andon KD-5905 Manuale operativo - Pagina 12
Sfoglia online o scarica il pdf Manuale operativo per Monitor della pressione sanguigna Andon KD-5905. Andon KD-5905 16. Fully automatic arm cuff blood pressure monitor
DN:KD-5905-SMSY01 V1.0
13. TROUBLESHOOTING (2)
PROBLEM
LCD shows low battery
symbol
LCD shows "Er 0"
LCD shows "Er 1"
LCD shows "Er 2"
LCD shows "Er 3"
LCD shows "Er 4"
LCD shows "Er 5"
LCD shows "Er 6"
LCD shows "Er 7"
LCD twinkling all
LCD shows "Er 9"
LCD shows "Er A"
LCD shows "Er b"
No response when you
press button or load
battery.
MAINTENANCE
1.
Do not drop this monitor or subject it to strong impact.
2.
Avoid high temperature and solarization. Do not immerse the monitor in water as this
will result in damage to the monitor.
3.
If this monitor is stored near freezing, allow it to acclimate to room temperature before
use.
4.
Do not attempt to disassemble this monitor.
5.
If you do not use the monitor for a long time, please remove the batteries.
6.
It is recommended the performance should be checked every 2 years or after repair.
Please contact the service center.
7.
Clean the monitor with a dry, soft cloth or a soft cloth squeezed well after moistened with
water, diluted disinfectant alcohol, or diluted detergent.
8.
No component can be maintained by user in the monitor. The circuit diagrams,
component part lists, descriptions, calibration instructions, or other information which will
assist the user's appropriately qualified technical personnel to repair those parts of
equipment which are designated repairably can be supplied.
9.
The monitor can maintain the safety and performance characteristics for a minimum of
10,000 measurements or three years, and the cuff can maintain the performance
characteristics for a minimum of 1000 measurements.
POSSIBLE CAUSE
Low Battery
Pressure system is unstable
before measurement
Fail to detect systolic pressure
Fail to detect diastolic pressure
Pneumatic system blocked or cuff
is too tight during inflation
Pneumatic system leakage or
cuff is too loose during inflation
Cuff pressure above 300mmHg
More than 3 minutes with cuff
pressure above 15 mmHg
EEPROM accessing error
Device parameter checking error
MCU self-verify error
Pressure sensor parameter error
EEPROM backup error
Incorrect operation or strong
electromagnetic interference.
12 / 16
Operation Guide
SOLUTION
Change the batteries
Don't move and try again.
Apply the cuff correctly and
try again
Measure again after five
minutes. If the monitor is
still abnormal, please
contact the local distributor
or the factory.
Take out batteries for five
minutes, and then reinstall
all batteries.
Template version:1