Changing automatic switch-off time
In order to change the automatic switch-off time, it is necessary to open a secure menu.
Open the menu only when you really have to. Accidental or
intentional changes to the parameters can have a significant impact
on the functionality of the scale.
To enter the secure menu, press and hold down the
switch on the scale and after it has switched on.
The parameter menu has 12 functions. The display shows the names of functions ("CAL",
"rES", "CAP", "Cnt", "GrA", "CoM", "bAUd", "oFF", "SPEEd", "MtArE", "rESt" and "bACk"). You
can navigate through these functions by repeatedly pressing the
press the
key. You can navigate through the values of this function by repeatedly
pressing the
key. To save your selection, press the
The menu can be exited by selecting the "bACk" parameter and pressing the
Caution: On calibrated scales, various functions such as "CAL" and "GrA" are locked. A
change leads to the loss of the calibration and is only reserved for qualified personnel
authorised by the manufacturer.
Indicator
Description
Setting the automatic switch-off time. The automatic switch-off time can be
set to 3, 5, 15 or 30 minutes ("oF 3", etc.) or the automatic switch-off function
oFF
can be disabled ("oF 0"). Press the
value. Press the
confirm the displayed value.
Cleaning
Disconnect the mains plug before cleaning the scale. Use only a damp cloth or an ordinary
disinfectant for cleaning. Do not use aggressive liquid cleaning agents, abrasive or acidic
detergents or white spirit.
Make sure that no liquid cleaning agent or water penetrates the scale and always follow the
manufacturer's instructions for use.