Askle Sante Winncare Automorpho Plus Manuale d'uso - Pagina 9
Sfoglia online o scarica il pdf Manuale d'uso per Apparecchiature mediche Askle Sante Winncare Automorpho Plus. Askle Sante Winncare Automorpho Plus 20.
WINNCARE FRANCE
Use:
Shows that the compressor is switched on
Flashes to indicate inflation of the support
After it goes off, the patient can be installed
The on LED shows that the keyboard is locked: it is not possible to
change mode.
Continuously pressing for 4 seconds unlocks the keyboard.
Locking is automatic after 5 minutes or it can be controlled by
continuously pressing for 4 seconds.
Low pressure dynamic mode: triggers the alternation of one cell
out of two (except for 2 cells for static head).
Treatment mode: makes the support surface firm in order to
facilitate handling of patient during treatment or lifting (e.g. Bed-
chair transfer).
The duration of this mode is limited to 30 minutes.
Securization of treatment mode: the flashing of the treatment
mode LED means that the mode is almost finished. It is triggered 5
minutes before the end. A sound signal is emitted when it is
triggered.
At the end of 30 minutes, the compressor automatically switches
to the previously used mode.
Low pressure static mode: inflates all cells in the support at a
constant pressure dependent on the morphology of the patient
and encourages its immersion in the support to increase the
support surface and thereby reduce average interface pressure.
Enables comfort to be adjusted for the patient. It is applicable in
the 3 modes of operation (DYNAMIC, STATIC and TREATMENT) and
in seated position.
By default, the LED is in the central position, which is the
calculated nominal position.
The seated position function can be selected when the chest
relief is over 30°.
When the function is set, dynamic cycles are accelerated for
greater comfort and restricts the risk of injury.
NOTE: It is preferable to press this function before tilting the chest
relief rather than the other way round.
'
NOTE: Don
t forget to come out of the function if the patient
returns to the lying down position.
FIXED LED: Compressor revision recommended
(17500 hours of operation)
Fixed LED: Low priority alarm. (See alarms chapter)
Flashing LED: Medium priority alarm.
(See alarms chapter)
- 8 -