- ページ 4

セキュリティセンサー Andor Technology AD02のPDF クイック起動マニュアルをオンラインで閲覧またはダウンロードできます。Andor Technology AD02 4 ページ。

1) Before connecting the USB cable, always power up the camera using the PSU supplied.
2) The camera has internal hardware protection from overheating. Should the body of the camera reach
>40º C, a buzzer will sound and the power to the cooler will be disconnected until a safe temperature is
reached.
3) For certain cameras, using a PS-25 power supply in Deep Cooling mode is designed for water cooling
only. It may, however, be possible to use this PS-25 power supply with air cooling depending on the
ambient air temperature. Typically the ambient air temperature must remain below 20 ˚C to dissipate the
extra heat generated in this mode of operation.
4) When a safe temperature is reached, the camera will Auto-Reset. If the buzzer sounds, check for the
following:
 Any obstructions in the air or water flow(s)
 Ambient air temperature & water temperature
 The temperature set in the Temperature screen. Set a lower temperature if required
Should you have any problems or queries, please contact the nearest Technical Support Engineer as
shown below:
International Office
Phone: +44 28 9023 7126
Fax: +44 28 9031 0792
Phone: +81 (3) 3518 6488
Fax: +81 (3) 3518 6489
QUICK LAUNCH GUIDE FOR
Need more information? Please contact us at:
Japan Office
NOTES
US Office
Phone: +1 (860) 290 9211
Fax: +1 (860) 290 9566
China Office
Phone: +86 (10) 5129 4977
Fax: +86 (10) 6445 5401
Page 4 of 4
CAMERAS