13.Data transmission settings
You can change the settings and the output data via USB.
■ Data Transmission Protocol
Baud rate: 19200
Character length: 8 bits
Stop bit: 1 bit
Parity bit: None
Delimiter: CR + LF
Flow control: None
■Transmission and reception code list
"Command"
"STAT"
Continuous transmission
"STOP"
Continuous transmission stop
"MEMORY"
Transmit all stored data
"CLEAR"
ALL CLEAR
"FUNC_MODE"
FUNC Switching
"TRACK"
Change to TRACK
"PEAK"
Change to PEAK
"PP"
Change to PP
"GF"
Change to gram unit system
"NEWTON"
Change to Newton unit system
"POND"
Change to pound unit system
"CMP_HI:"
Comparator HI setting
"CMP_LO:"
Comparator LO setting
"OVER:"
Overload setting value
"LCD_LATE:"
Screen update rate
"LCD_FLIPPED:"
Flip the screen vertically
"POWER_OFF:"
Auto power off
"AUTO_CLEAR:"
ALL CLEAR
"MIN_DIGIT:"
Minimum digit display
"BUZZ:"
Buzzer permission
"BACK_LIGHT:"
Backlight permission
"GAIN_INVERT:"
Plus / minus inversion
"INIT_VALUE:"
Reset to initial setting
"PRINTER:"
Printer permission
"SET_SAVE"
Save change value
"CANCEL"
Cancel the change
"VALUE_READ"
Read current value
* An example of send command input
Set the comparator HI to 10.0.
Command: setting value CR + LF
COMP_HI: 10.0 CR+LF
* Receive a reply message of "COM_OK!!" if the command input is correct.
* Receive a reply message of "COM_ERR!!" if the command input is incorrect.
* Be sure to send "SET_SAVE" after changing the settings. Otherwise, the settings will not be enabled and the display
will not return to the measurement screen, either.
* Once set, it cannot be returned to the original value.
Contents
The signal is transmitted at 2KHz.
(The output continues until the "STOP" signal is received.)
Delete the stored data
Please send first when you change the settings.
Error if more than the settable value.
Cannot exceed the maximum value.
Error if more than the settable value.
Cannot exceed the comparator HI setting.
Error if not a settable value.10~120 in 10 increments OVER:100
Error if not a settable value.20,10,5,2 only
"ON": permission, "OFF": none
"ON": permission, "OFF": none
"ON": permission, "OFF": none
"ON": permission, "OFF": none
"ON": permission, "OFF": none
"ON": permission, "OFF": none
"ON": permission, "OFF": none
"ON": permission, "OFF": none
"ON": permission, "OFF": none
Send it at last to save values.
Send it at last to cancel the change.
Read current value. COM_OK is not returned.
Cautions
Example (all with CR+LF attached at the end)
STAT
STOP
MEMORY
CLEAR
FUNC_MODE
TRACK
PEAK
PP
GF
NEWTON
POND
CMP_HI:2000
CMP_LO:1000
LCD_LATE:2
LCD_FLIPPED:ON
POWER_OFF:ON
AUTO_CLEAR:ON
MIN_DIGIT:OFF
BUZZ:OFF
BACK_LIGHT:ON
GAIN_INVERT:OFF
INIT_VALUE:
PRINTER:ON
SET_SAVE
CANCEL
VALUE_READ