Evaluation of the Body Mass Index

Compare the determined value which corresponds to those being used by the WHO.
BMI below 18,5
The patient is under the regular standard and does not got enough weight. An increase of
weight is recommended. Please consult a professional.
BMI between 18,5 and 24,9
The patient got a normal weight. If he feels well there is no need for a change.
BMI between 25 and 30
The patient is slightly overweighed to pre-obese. A reduction of weight is indicated or, is
necessary if the state of health is already affected (e.g. diabetes, high end blood pressure,
heart diseases).
BMI over 30
Reduction of weight is imperative. To visit a doctor is highly recommended.

Connecting additional devices

Connection of printer, PC or remote-display will be done with the RS-232 interface. For connection of
additional devices kindly ask your local representative or authorized ADE dealer for further instructions.

Print out a record

The scales must be at standstill. The standstill status symbol
ENTER key to send the data to the serial interface.
Note: Printing is not possible if the net weight is 0. The print function only produces one print-out. A reset
is required to print out another copy.

Error Messages

Error Codes
Load cell wires to indicator are
incorrectly connected; or shorted, or
opened; or ADC/ load cell are being
damaged
Weight reading exceeds the zero limit
after switching the scale on.
Weight reading is below the zero limit
after switching the scale on.
Weight reading exceeds load limit.
Weight reading below load limit.
Configuration or Calibration parameters
are not correctly set.
7
Description
must appear at the screen. Press the
Solution
Make sure wires are in good order
and condition and correctly
connected. Replace load cell or
ADC chip. Service is required.
Make sure scale platform is empty.
Install platform on scale.
Otherwise the zero calibration/
adjustment have to be done by
authorized personnel.
Reduce load on scale until weight
value can be displayed.
Service by authorized personnel
required.
Contact your local dealer / supplier
for calibration purpose.