送信機 AFRISO 31836のPDF 取扱説明書をオンラインで閲覧またはダウンロードできます。AFRISO 31836 9 ページ。 Differential pressure transmitter

DMU 10 / DMU 11

2. Product identification

The device can be identified by its type plate. It provides the most important data. By the or-
dering code, the product can be clearly identified.
designation
Diff. Pressure range
Signal
Suppy
Fig. 1: type plate DMU 11 D
3. Installation
3.1 Assembly information
• Keep in mind that this device is an electronic precision measuring device. Handle the de-vice
carefully and properly to avoid any damages!
• Do not throw the device!
• Remove the packaging and, if necessary, the protective cap of the device only shortly before
installation in order to prevent damage to the diaphragm!
• Keep the protective cap!
• After disassembling, reattach this protective cap over the diaphragm.
• Handle an unprotected membrane with extreme care; This can be easily damaged.
• Do not use any force when installing the devices!
• Make sure that no mechanical stresses occur at the pressure port with low pressure ranges
during installation, as this may cause a shifting of the characteristic curve..
• For the connection of the pressure lines, a sealing has to be installed by the operator.
• For the pipe assembly, an installation as zero-potential as possible must be observed.
• Consider for the installation of a DMU 10 D, that the pressure ports must not be turned
against the housing!
3.2 General installation steps
1. Remove the pressure measuring device from the package carefully.
2. Connect the reference pressures according to the following installation steps
and to your mechanical connections. Therefore, you will have to observe that:
- The higher pressure must be connected with the input "+" (DMU 10 D) or
"P1" (DMU 11 D)
- The lower pressure must be connected with the input "-" (DMU 10 D) or
"P2" (DMU 11 D)
Typ
Ordering
code
Serial
number
Connector
pinout
3