Daily Verification-Step One
Primary Standard
1
Locate a vial of the 0.760 Water Activity Standard that
is inside the carrying case with your instrument or in a
box purchased separately (Blue).
Note: Note: Each Pawkit case contains a set of trial vials. You will need to
order more standards for your daily verification of the instrument.
Twist off cap of the 0.760 vial and pour the contents
2
into a sample cup.
Place the Pawkit over the sample cup and turn on the
3
instrument by pushing the left button. If the instrument
is already on just press the left I button again to start
another reading.
The lids with the cups are used to prevent
moisture loss or gain-if you do not immediately
Important!
test samples. Do not place a sample cup with a
lid under the Pawkit.
4