When using IX83/IX73
1
Loosen the U-LHEAD clamping screws a , and separate the U-LHEAD to 2 units: one for the frame and the other for
the lamp housing. One for the frame displays [FRAME] and the other for the lamp housing displays [
2
Loosen the lamp housing clamping screws b (2 pcs.) on the fluorescence illuminator sufficiently using the Allen
screwdriver, fit the mount dovetail c of the U-LHEAD (frame side) and tighten the clamping screws b .
3
Attach the U-AW d to the U-LHEAD (frame side), and tighten the clamping screws a .
4
Loosen the clamping screws e (4 pcs.) of the U-AW, fit the U-LHEAD (lamp housing side) f and tighten the clamping
screws e (upper 2 pcs. only).
5
Attach the lamp housing g to the U-LHEAD (lamp housing side) f in the same way as above.
6
Insert one end of the interface cable into the connector h and insert the other end into the control box IX3-CBH or
control box IX3-CBM (see "SYSTEM CHART").
d
a
c
b
g
h
f
e
U-AW
].
7