- ページ 9
ポータブル発電機 ACD CAL-101のPDF 取扱説明書をオンラインで閲覧またはダウンロードできます。ACD CAL-101 15 ページ。 Calibration gas generator
If the cell has been depleted, it should be removed and replaced with
a new generating cell. Please refer to the section on cell life for proper
instructions on replacing the generating cells. If the cell has not been
depleted, there may be a faulty electrical connection between the cell and
the instrument. Remove the cell and inspect the pins protruding from the
pump board and the corresponding pads on the cell board. Remove any
debris and or corrosion that may be present and reinsert the cell ensuring
proper alignment. If either the pins become damaged or non-repairable,
the pump board should be replaced..
The rate of airflow through the instrument is critical for accurate gas
concentrations. The flow should be adjusted to 0.5 LPM (liters per minute)
STP (standard temperature and pressure) of sea level and 20
C. (see
o
chart in section VI) To trim the flow hold the magnetic screw driver over
the 'flow trim' area of the instrument indicated with a magnet symbol. The
GREEN LEDs will begin flashing at a once per second rate indicating the
relative power to the pump. As you hold the magnet over this area the pump
will continue to increase in speed (The LEDs will ramp up indicating the
relative power) until the magnet is removed. When the magnet is reapplied
to the area, the pump speed will decrease (the LEDs will also ramp down
indicating the relative power) until the magnet is again removed. Each time
the magnet is removed and reapplied the flow will change from increasing to
decreasing. This is how the flow rate of the instrument should be adjusted
to obtain 0.5LPM STP. Once the flow is adjusted the instrument will save
the flow data and return to that flow rate for future use
Page 7