Removing the power supply and controls
assembly
1.
Rest the instrument on its back side.
2.
Remove the two screws [B] that secure the power
supply and controls assembly to the bottom of the
chassis.
B
3.
Place the instrument in the upright position.
4.
Remove the two screws that secure the power
supply and controls assembly to the rear of the
chassis.
5.
Slide the power supply and controls assembly out
of the base assembly.
OsmoTECH® PRO Sample Probe Motor
Removing the osmometer module
1.
Remove the four socket head cap screws and
washers [D] that mount the osmometer module
mounting bracket to the chassis.
2.
Carefully slide the module out of the instrument.
3.
Remove the hex nut and washer [E] that secures
the SPG and BPG ring terminals to the side of the
osmometer module mounting bracket.
4.
Remove the four socket head cap screws and
washers [F] that mount the osmometer module to
the osmometer module mounting bracket (which
is secured to the chassis).
F
5.
Lift the osmometer module off of the osmometer
module mounting bracket.
Replacement Procedure
D
F
F
E