Installation (cont)
Sensor Testing
Occupancy Sensors
Ensure the bed/chair is unoccupied and after pressing RESET there will be a four
tone beep and the unit should be in STANDBY. Five tones will indicate the sensor
is being falsely triggered by the weight of the bedding/cushion or otherwise in-
stalled incorrectly.
Activate the sensor and a single beep should be heard indicate successful detection
of the occupant. After that operation is silent.
Deactivating the sensor will then raise the alarm, either instantly or after the delay
period (it is not essential to wait for this as the above proves the senor operation).
Reset the unit to stop further nuisance alarms
Switch Sensors
After RESET there should be no alarm. Activate the sensor and check an alarm is
raised. There is an optional beep when the sensor is activated (i.e. usually for
switch inputs but not for floor mats).
These tests should be performed regularly to confirm correct system operation
All parts of the system, apart from the pager charger, are suitable for use within the patient vicinity
Operation
The monitor is turned on by
Ÿ Removing the power key for battery power
Ÿ Connecting a mains adapter if required
Ÿ Or its is always available if the rechargeable battery has
been fitted and enabled
Operating the P163:
Call for Assistance (all versions)
Activating the call button for longer than 1 second will result
the Nurse Call being activated.
P163BAB Bed/Chair Vacation
After power-up (or RESET) the monitor will be in STANDBY. When the
Occupancy sensor is activated there will be a beep (default) and the monitor will
become active. When the user leaves the bed/chair, the monitor will activate the
Nurse Call (either immediately or after a time delay dependent on the "1-Input
UH1142B P163 Wired General Interface v2 page 4 of 8