3.3 Wiring of the junction box

1. Unscrew the four screws (9) on the cover (8) of the junction box (7), remove the cover (8) and the gasket (10).
2. Remove the end cap (11) from cable glands before passing cables throw cable glands M20 (3,6).
The cable glands M20x1.5 can be used for cables of a diameter of 7-13mm.
3. Pass the power/data cables through the cable glands (3,6) and connect them to the connection blocks (12, 13).
4. Place the See their gasket (10) on the junction box (7) and screw the cover (8) to the junction box (7).
Check that all cable glands (and blind flange) and screws are firmly secured.
4. After checking all connections, fill the junction box with synthetic resin. Make sure that the End cap is inserted
in unused cable gland before filling it.
At filling of the junction box proceed according instructions stated on the bag of the resin.
The connection blocks have to be fully poured in the resin.
Wiring of connection blocks (12,13) in the junction box.
CE version
Connector
Function
Colour of wire
+
D+
LEDs +
Data +
Red
Orange
Eminere Remote UV
D-
Data -
Black
8
-
LEDs -
Ground
White
Not connected