ADE EHR1 운영 매뉴얼 - 페이지 13

{카테고리_이름} ADE EHR1에 대한 운영 매뉴얼을 온라인으로 검색하거나 PDF를 다운로드하세요. ADE EHR1 20 페이지. Platform scale

ADE EHR1 운영 매뉴얼
P12.x
0/1/2/3/
4/5/6/7/
8/9
P13.x
0/1/2
P14.x
0/1/2/3
P15.x
0/1/2/3/
4/5/6/7/
8
P16.x
0/1/2/3
P17.x
0/1/2/3/
4/5/6/7/
8/9
P18.x
0/1/2/3/
4/5/6/7/
8/9
Defaults are marked in bold type. Changes to the technical settings may cause the
balance to malfunction and void the warranty.

Cleaning

Before cleaning, disconnect the device from operating voltage. Do not use acid cleaning
agents but only a dump cloth or common disinfection product for cleaning.
Pay attention that no fluid penetrates the scale!
EHR_EHQ-210517-Rev009-UM-en
Zero range for ZERO button
(Modifications in setting only by authorized personsl!)
Power-on zero-point 0: +/-1%FS, 1: +/-2%FS, 2: +/-3%FS,
3: +/-4%FS, 4: +/-5%FS, 5: +/-10%FS, 6: +/-20%FS,
7: +/-50%FS, 8: +/-100%FS, 9: no limitation
Weight signal within power-on zero point range, change
which data as current power-on zero point
(Modifications in setting only by authorized personsl!)
0: current weight
1: calibration zero-point
2: switch-off zero-point
Weight signal not within power-on zero point range, change
which data as current power-on zero point
(Modifications in setting only by authorized personsl!)
0: current weight
1: calibration zero-point
2: switch-off zero-point
3: continuously display „0 - - - -"
Zero tracking range
(Modifications in setting only by authorized personsl!)
0: 0d = no tracking, 1: +/-0,25d, 2: +/-0,5d, 3: +/-1d,
4: +/-1,5d, 5: +/-2d, 6: +/-3d, 7: +/-4d, 8: +/-5d
Data filter intensity
0: very weak, 1: weak, 2: middle, 3: strong
Check weight stability range
0: +/-0,5d, 1: +/-1d, 2: +/-1,5d, 3: +/-2d, 4: +/-3d, 5: +/-4d,
6: +/-5d, 7: +/-6d, 8: +/-7d, 9: +/-8d
Overload limit range
0: FS+0d, 1: FS+9d, 2: 101%FS, 3: 102%FS, 4: 105%FS,
5: 110%FS, 6: 120%FS, 7: 150%FS, 8: 120%FS,
9: no limitation

Cleaning and Maintenance

12