ADE M601020 운영 매뉴얼 - 페이지 7
{카테고리_이름} ADE M601020에 대한 운영 매뉴얼을 온라인으로 검색하거나 PDF를 다운로드하세요. ADE M601020 16 페이지. Class iii approved scales electronic bed scale
Menus structure of setting mode:
After entering the clear button following options are available and appear on the screen when moving with the
navigation buttons:
FILTER
SERIAL
Setting date and time
–
Enter the CLEAR key for approx. 4 sec; FILTER appears on the display.
–
Press key "6" six times and TIME appears on the display.
–
Press ENTER key and SHOW appears on the display.
–
Press key "6" one time and HOUR appears on the display.
–
Press ENTER key and set the hours using the arrow keys "2", "4", "6" and "8".
–
Press the ENTER key and set the minutes and seconds.
–
Go back by pressing key "2"; TIME appears on the display.
–
Press key "6" one time and DATE appears on the display like the time entry.
–
Press key "2" two times and EXIT Y appears on the display.
–
Confirm with ENTER and date & time are set.
Modes of operation
Weighing mode
The "weighing mode" is of course the standard mode of your scale. After initial configuration and sealing (by
attaching a calibration sticker to the back of the unit), the ADE-scale can only be operated in this mode.
Setting mode
Date, time and consecutive numbers for print commands are being set in this mode. To start setting mode,
hold the GROSS/NET key pressed until TIME appears on the display.
Configuration mode (set-up mode)
The values for different parameters including the calibration could be set in the configuration mode. This func-
tion is for authorized service personnel only.
Test mode
There are different diagnostic functions for the ADE-MAK in test mode. This is for service personnel only.
Manipulations to test and configuration modes can revoke the calibration status and will result
in loss of warranty!
Before weighing
1. Move the wheels of the bed onto the weighing platforms.
2. The constructions do not allow to pass over the weighing platform. In any case ensure that the wheels of
the bed are secured and the brakes are engaged.
3. Switch on the scale.
4. An automatic function test will lead to zero indication (0.0). The weight of the bed is being ignored auto-
matically.
5. Place the patient into the bed. The displayed weight is the weight of the patient.
6. If the weight of the patient should be ignored as well (e.g. for dialysis purposes), follow the instruction un-
der the point "Tare function".
7. If the weight of the bed including bed-clothes is already being stored follow the instruction under the point
"Restore the preset tara value"
M601020-210322-Rev009-UM-e
PROGR
PFORMT
SETPTS
TEST
TIME
Operation
DATE
VERS
7